THE MEASUREMENT OF ENVIRONMENTAL RADIATION EXPOSURE AROUND THE LINAC RADIOTHERAPY BUNKER
Authors
Septiana Kurniasari , Fatimah K Hentihu , Alfian K Anto , Henry PrasetyoDOI:
10.29303/ipr.v5i1.129Published:
2022-01-31Issue:
Vol. 5 No. 1 (2022)Keywords:
radiation exposure, linac, radiotherapyArticles
Downloads
How to Cite
Abstract
Radiotherapy is cancer therapy using radiation. One of the tools used for radiotherapy is the Linac. Using Linac also can have a detrimental effect on radiation workers, patients, and the public. The reaction can cause activation around the room to become radioactive. Therefore, radiation protection is needed. This research aimed to measure environmental radiation exposure as a radiation protection effort around the Linac bunker. This research uses a quantitative method by collecting data directly on the area around the Linac bunker, with a distance of ± 30 cm from the bunker wall, using the STEP OD-02 Detector. The average Linac operation is 2000 hours/year, while the safe dose limit used is 1/2 dose Limit from Nuclear Power Monitoring Agency. Based on measurements of environmental radiation exposure, there are eight areas classified as safe and four other regions experiencing radiation leakage, including Doors with a radiation exposure difference of 0.078 Sv/hour; West Field (0.074 Sv/hour); TPS Room (0.302 Sv/hour); and Meeting Rooms (0.199 Sv/hour), but still within the safe limit of radiation dose. The environmental radiation exposure around the Linac Radiotherapy bunker is safe.References
N. Fitriatuzzakiyyah, R. K. Sinuraya, I. M. Puspitasari. (2017). Radiation Cancer Therapy: Basic Concepts of Radiotherapy and Its Development in Indonesia. Indonesian Journal of Clinical Pharmacy, 6(4), 311-320.
R. A. Puspitasari, W. I. Pertiwi, P. M. Sholihah, W. H. Fariqoh, N. Kavilani, S. D. Astuti. (2020). LINAC Radiation Beam Quality Analysis for Radiotherapy Effectiveness. Postgraduate Journal of Bioscience, 22(1), 11-19.
E. Guritna, G. Maslebu, N. A. Wibowo, M. Hidayatullah. (2017). Electron Beam Profile Constancy Analysis on Linac. FLUX Physics Journal, 14(2), 110-119.
M. Vadila & D. Milvita. (2018). Electron Beam Output Analysis of LINAC Therapy Variant Type CX 6264 at Andalas University Hospital. Journal of Physics Unand, 7(2), 91-96.
K. N. Widyanti, C. S. Widodo, Bunawas. (2018). Study of Thermal Neutron Flux Distribution and Energy Produced by Linac Using Detector CR-39 in Water Medium. NATURAL B, 4(3), 145-152.
Mayarani, E. P. S. Hidayat, N. H. Apriantoro, R. Kristian, M. Irsal. (2018). Analysis of Rooftop Skyshine Radiation Exposure with Angle of Gantry Linear Accelerator 180° in Radiotherapy Unit of Pertamina Central Hospital Jakarta. SANITAS: Journal of Health Technology and Arts, 9(1), 24-34.
B. S. W. Hadi& D. Milvita. (2018). Verification of Field Area Radiation Radiation Clinac CX Type Integrated Electronic Portal Imaging Device (EPID) Using IMRT Technique at Andalas University Hospital. Journal of Physics Unand, 7(4), 334-338.
D. Milvita &B. S. W. Hadi. (2019). Verification of Geometry and Gamma Index on the EPID Integrated Clinac CX Type Linac at Andalas University Hospital. Wahana Fisika, 4(2), 111-119.
D. N. Dasril & O. P. Sari. (2020). External Dose Measurement Received by Radiography Engineering Practicum Students at the Radiology Laboratory, Baiturrahmah University, Padang. Journal of Theory and Application of Physics, 8(1), 35-42.
M. Utari, D. Milvita, N. Nuraeni, H. Yuliati. (2014). Radiation Dose Analysis for Radiotherapists Using Pocket Dosemeter, TLD Badge and TLD-100 at the Radiotherapy Installation, RSUP dr. M. Djamil Padang Case Study (May – October) 2014. Journal of Physics Unand, 3(4), 262-268.
D. Rochmayanti, S. Daryati, Darmini, Y. Kartikasari. Profile of Radiation Exposure of Radiology Installations in an Effort to Support Protection Programs at Hospitals/Radiology Clinic Laboratories in the City of Semarang. JImeD, 5(1), 20-24.
Regulation of the Nuclear Power Monitoring Agency of the Republic of Indonesia Number 4 of 2020 concerning Radiation Safety in the Use of X-Ray in Diagnostic and Interventional Radiology.
Rudi, Pratiwi, Susilo. (2012). The Measurement of X-Ray Radiation Exposure in Radiodiagnostic Installations for Radiation Protection. Unnes Physics Journal, 1(1), 19-24.
Nazaroh, S. Muhammad, G. Wurdiyanto. (2016). Application of Radiation Protection and Safety at the Radiotherapy Facility-Linac. Proceedings of the Nuclear Safety Seminar, 331-339.
R. Akhbar, G. Anindita, M. Y. Santoso. Redesign of Linear Accelerator (Linac) Shielding Room for Radiation Safety in Building 14 PSTA-BATAN Yogyakarta. Proceeding 1st Conference on Safety Engineering and Its Application, ISSN No. 2581-1770.
Author Biography
Septiana Kurniasari, Gorontalo State University
License
Authors who publish with Indonesian Physical Review Journal, agree to the following terms:
- Authors retain copyright and grant the journal right of first publication with the work simultaneously licensed under a Creative Commons Attribution-ShareAlike 4.0 International Licence (CC BY SA-4.0). This license allows authors to use all articles, data sets, graphics, and appendices in data mining applications, search engines, web sites, blogs, and other platforms by providing an appropriate reference. The journal allows the author(s) to hold the copyright without restrictions and will retain publishing rights without restrictions.
- Authors are able to enter into separate, additional contractual arrangements for the non-exclusive distribution of the journal's published version of the work (e.g., post it to an institutional repository or publish it in a book), with an acknowledgment of its initial publication in Indonesian Physical Review Journal.
- Authors are permitted and encouraged to post their work online (e.g., in institutional repositories or on their website) prior to and during the submission process, as it can lead to productive exchanges, as well as earlier and greater citation of published work (See The Effect of Open Access).